SUPPORT
Tissues lost or tissue morphology is unclear

Possible reason:Cells are adhered together

Recommended solution:Shorten the digesting time


Possible reason:Cells are seriously fragmented

Recommended solution:Too many necrotic areas in the sample, the section thickness is inaccurate, and the time for the digestion step is too long 


Possible reason:The outlines of cells are not clear, or cloud-like matter covers the cells 

Recommended solution:The time for the digestion step is too short


Possible reason:Improper glass slides are used

Recommended solution:Use the positively charged glass slides 


Possible reason:The glass slide is baked incorrectly

Recommended solution:Check the temperature of the slide warmer


Possible reason:Pretreatment and denaturation are excessive 

Recommended solution:Shorten the pretreatment and denaturation time

Background is too strong

Possible reason:The glass slide is not cleaned before the sample is made

Recommended solution:Wash the glass slide with absolute ethanol


Possible reason:Washing is insufficient after hybridization

Recommended solution:Ensure that the PH value and temperature of the washing liquid are correct, remove the cover glass, and wash repeatedly


Possible reason:The filter group is used improperly

Recommended solution:Replace the appropriate filter group to observe to reduce the background light


Possible reason:The washing temperature is too low

Recommended solution:When the glass slide is washed, ensure that the temperature of solution reaches the temperature required in the LBP’s manual 


Possible reason:The sample is not well fixed

Recommended solution:Try to replace with other paraffin blocks for treatment, or recommend pathologists to communicate with clinicians in time for fixation time and effect


Possible reason:The washing liquid is used for a long time or stored incorrectly

Recommended solution:Ensure the use frequency of washing liquid, replace the washing liquid in time, and store the washing liquid according to the requirements of LBP’s manual

No signals or signals are weak

Possible reason:Samples are denatured insufficiently

Recommended solution:Ensure that the denaturation temperature of the hybridizer reaches the temperature required by the LBP’s manual, and the hospital needs to calibrate the equipment regularly


Possible reason:The probe is not sufficiently mixed before use

Recommended solution:Blow and beat the probe, mix well, and centrifuge shortly


Possible reason:The probe is not added

Recommended solution:Fully thaw the probe, to ensure that the pipette absorbs the reagent in the probe 


Possible reason:The probe is expired

Recommended solution:Use the probe within the validity period


Possible reason:The amount of the probe is insufficient

Recommended solution:Ensure that the pipette sucks accurately, and increase the amount of the probe to 10 ul. Please do not dilute the probe


Possible reason:Bubbles are formed under the cover glass during hybridization

Recommended solution:The cover glass should cover the surface of the probe and tissues when placed , and should be squeezed gently to squeeze out bubbles


Possible reason:Hybridization conditions are unsuitable

Recommended solution:Set according to the requirements of LBP’s manual


Possible reason:The temperature of the hybridizer about digesting,washing and overnight are incorrect

Recommended solution:Ensure that the equipment such as the water bath meets the temperature required in LBP’s manual


Possible reason:Washing liquid or washing conditions are incorrect

Recommended solution:Set according to the requirements of LBP’s manual


Possible reason:Sample glass slides or probes are stored incorrectly

Recommended solution:Set according to the requirements of LBP’s manual


Possible reason:Counterstain is used incorrectly or expired

Recommended solution:Use DAPI within the validity period, set and use DAPI according to the requirements of LBP’s manual


Possible reason:The filter group selected during observation is inappropriate

Recommended solution:Recommended to use the parameter settings of the reagent filter required in LBP’s manual


Possible reason:The sample is not well fixed

Recommended solution:Try to replace with other paraffin blocks for treatment, or recommend pathologists to communicate with clinicians in time for fixation time and effect


Possible reason:The front and back of the glass slide are error

Recommended solution:Retreat, and pay attention to the front and back when the glass slide is taken out


Possible reason:Glass slides are dried insufficiently

Recommended solution:Before the probe drops the reagent onto the glass slide, guarantee that the ethanol solution on the glass slide has been fully volatilized


Possible reason:The film-sealing gum or cover glass are not removed

Recommended solution:Remove all film-sealing gum in time and uncover the cover glass with moderate strength


Possible reason:Proper film-sealing gum is not used

Recommended solution:Use special Fish Fixogum film-sealing gum. Do not use sealing film, transparent adhesive top, and preservative film


Possible reason:The service life of microscope mercury lamp is expired

Recommended solution:Please contact the microscope manufacturer to replace with a new mercury lamp


Possible reason:The microscope structure and objective lens are not appropriate for observing FISH samples, or the filter is oxidized

Recommended solution:Please contact the microscope manufacturer

Tissues fall off from the glass slides during dyeing

Possible reason:Glass slides are not pretreated before fixation, resulting that samples are the adhered with glass slides

Solution:Use glass slides with adhesives

Backgrounds of all sections are dyed

Possible reason:Enzymatic activity of endogenous peroxides is not completely blocked

Solution:Block with 3% H2O2


Possible reason:Non-specific dyeing

Solution:Add 10% normal goat serum before adding antibody


Possible reason:The antigen is diffused due to autolysis of tissues

Solution:Use fresh tissues or fixed tissues in time as far as possible


Possible reason:Incomplete deparaffinating

Solution:Replace xylene and alcohol


Possible reason:Sections are flushed insufficiently

Solution:Flush completely


Possible reason:Enzyme reaction accelerates at too high room temperature

Solution:Ensure that the room temperature is 15-25 to shorten the developing time


Possible reason:Sections are dry during dyeing

Solution:Ensure that sections are moist

The glass slides to be tested are not dyed, but the positive control is dyed

Possible reason:The antigen is denatured or blocked in the fixing or embedding process

Solution:It is recommended to use 10% neutral buffered formalin fixative;Ensure correct fixation time and baking temperature


Possible reason:The antigen is decomposed due to autolysis of tissues

Solution:Fix the samples in time

12
ONLINE MESSAGE
Please enter your contact information and we will get back to you as soon as possible.
submit
×